Connecting The Bar Code Scanner - Bayer HealthCare RapidPoint 400 Series Service Manual

Table of Contents

Advertisement

Connecting the Bar Code Scanner

Materials:
bar code scanner kit, which includes the bar code scanner, cable, and
holder
power from the system as described in this procedure before you connect
the bar code scanner.
1. Remove power from the system:
a. If prompted, enter your password, the service password, or the
password of the day.
b. Touch the
c. Touch
Cartridges installed in the system remain stable for 60 minutes
CAUTION:
without power. To maintain cartridge stability, do not remove power from
the system for more than 60 minutes if a cartridge is installed.
d. When the Shutdown screen appears, turn the power switch off.
2. Connect the bar code cable to the bar code scanner connector.
The bar code scanner connector is located on the rear panel of the
system and is labeled with a bar code symbol.
3. Tighten the hold-down screws on the connector.
4. Attach the holder for the bar code scanner to the right side of the
system.
5. Turn the power switch on.
After the Rapidpoint 400 series system title screen appears, the Wait
screen displays the time remaining until you can use the system. The
Analysis screen appears when the system is ready to use.
looking directly into the light beam to prevent possible exposure to
hazardous light.
Rev. C
To prevent electrical shock or damage to the system, remove
button.
Status
.
Shutdown
The bar code scanner emits a low-power visible laser. Avoid
Connecting the Bar Code Scanner 409