Installing The New Rapidpoint 400 Series System - Bayer HealthCare RapidPoint 400 Series Service Manual

Table of Contents

Advertisement

12. Wipe the exterior surfaces of the Rapidpoint 400 series system with a
10% bleach solution (0.5% sodium hypochlorite).

Installing the New Rapidpoint 400 Series System

1. Unpack the shipping box and remove any packaging materials from
the system.
2. Place the system on a bench top or other level work surface.
The system must equilibrate to room temperature to successfully
complete the installation. If the ambient temperature is near the limits
of the recommended operating temperature (15 to 32°C), the touch
screen may become out of calibration.
3. Remove the tape from the outside of the Rapidpoint 400 series system.
4. Install a new roll of paper in the printer.
5. Connect the following to the back of the Rapidpoint 400 series system:
bar code scanner
power cord
Do not connect the RJ-45 network cable at this time.
NOTE:
6. Turn on the power switch and perform an emergency touch screen
calibration:
a. Carefully watch the screen as the system boots up.
b. When the Bayer Diagnostics screen appears, touch it only once
within 2 seconds.
c. After a few minutes of loading the software, the system prompts
you for the touch screen calibration.
d. When prompted, perform the touchscreen calibration by touching
the two target buttons.
e. The first button appears in the lower left-hand corner. The second
button appears in the upper right-hand corner.
414
Rev. C