a. While facing the valve alignment gauge, slide the guide at the end
of the gauge to the left.
b. Align the dowel pins on the valve alignment plate with the holes on
the back of the valve alignment gauge.
c. When the valve alignment plate is in place, release the guide.
Ignore the small gauge.
NOTE:
d. While holding the valve alignment plate against the valve
alignment gauge, pull up and verify that the gauge reads
within 0 ± 0.5.
If the result . . .
is not within the acceptable
range
is within the acceptable range
Before removing the measurement cartridge, ensure that any
CAUTION:
tests not requiring removal of the measurement cartridge are performed.
Once a measurement cartridge is removed, if reinstalled it is likely to fail
within a few days. To verify repairs, you may reinstall the measurement
cartridge while remaining in Diagnostics. A new cartridge must be installed
before leaving the customer site.
Rev. C
Then . . .
loosen the set nut on the gauge and adjust
the gauge to zero. When adjustment is
complete, tighten the set nut.
Go to step 2.
go to step 2.