Error Messages And Remedies - Bayer HealthCare CLINITEK 50 User Manual

Table of Contents

Advertisement

Error Message
REPLACE BATTER IES
STR IP PROBLEM - RETEST
HELP
TABLE PROBLEM - RETEST
HELP
INSTRUMENT ERROR X
HELP
Possible Cause
and Remedy
The batteries do not have enough
power to test any more strips. Turn the
power off and replace the batteries.
A test strip is not present, is dry, or
is not positioned properly (upside
CONT
down, not touching the end of the
test table, or not laying flat in the
middle trough); the wrong table is
inserted for the strip being used;
or the test being used cannot be
identified by the software. Discard
the strip, then press CONT and
retest using a new test strip and the
correct table; be sure the name of
the strip you are using is shown on
the display.
The test table is not in its correct
position. The instrument or table
CONT
may have been moved or bumped
during testing, or the bottom of the
table may be dirty. Discard the strip.
Remove and clean the table, then
reinsert. Press CONT and retest
with a new test strip.
Error 3: The table is not present or is
not pushed in far enough to allow
CONT
movement. Push the table in about
halfway.
All other Errors: Discard the strip.
Turn the instrument off, wait several
seconds, then turn it back on. If the
error occurs again, contact Customer
Service (see page 25).