Calibrating The Touch Screen - Bayer HealthCare CLINITEK 500 Operating Manual

Urine chemistry analyzer
Hide thumbs Also See for CLINITEK 500:
Table of Contents

Advertisement

5.
c. Place the printer fully into position by lowering the
front of the module down and forward under the
ledge. Then lower the back of the module and press
down firmly until it snaps into place.
5.
d. Replace the clear shield to its position in front and
on top of the printer and press gently against the
adhesive pads.
5.
e. Replace the roll of paper, as instructed in Section
5, "Changing the Paper."
5.
f. Plug in and turn on the CLINITEK 500 instrument.
Test the new printer by printing the instrument
setup parameters (from the Ready/Run screen,
touch Menu, then Print, then Setup report) or by
performing several Reagent Strip tests. (If you had
turned the internal printer off prior to replacement
of the module, be sure to turn it back on through
the Setup Routine.)
5. f.
NOTE: If "
Printer Error
attempting to print, check for a tight and proper
connection of the interface cable.
Revised 10/06

Calibrating the Touch Screen

a key is touched, the screen may need to be calibrated.
This is easily done, as follows:
1. Turn the instrument power off. Wait several seconds,
2. When the title screen is displayed (which contains the
3. The display will prompt "
" is displayed when
If the touch screen does not respond correctly when
then turn the power back on.
software version and copyright information), touch
the screen anywhere.
Touch the top left cor-
" and an "X" will be displayed in one corner. Touch
ner
the screen, directly on top of the "X". Repeat when
the prompt changes to "
Touch the bottom right
" When the touch screen has been recali-
corner.
brated, the display will automatically continue in the
normal sequence of screens.