Why Can't The Dmm6500 Read My Usb Flash Drive; How Do I Change The Command Set - Keithley DMM6500 User Manual

6 1/2-digit bench/system multimeter
Hide thumbs Also See for DMM6500:
Table of Contents

Advertisement

DMM6500 6½ Digit Multimeter User's Manual

Why can't the DMM6500 read my USB flash drive?

Verify that the flash drive is formatted with the FAT32 file system. The DMM6500 only supports FAT
and FAT32 drives using MBR (Master Boot Record).
In Microsoft
drive.

How do I change the command set?

You can change the command set that you use with the DMM6500. The remote command sets that
are available include:
SCPI: An instrument-specific language built on the SCPI standard.
TSP: A scripting programming language that contains instrument-specific control commands that
can be executed from a stand-alone instrument. You can use TSP to send individual commands
or use it to combine commands into scripts.
SCPI2000: An instrument-specific language that allows you to run code developed for Keithley
Instruments Series 2000 instruments.
SCPI34401: An instrument-specific language that allows you to run code developed for Keysight
Model 34401 instruments.
If you change the command set, reboot the instrument.
You cannot combine the command sets.
As delivered from Keithley Instruments, the DMM6500 is set to work with the DMM6500 SCPI
command set.
If you choose the SCPI2000 or SCPI34401 command set, you will not have access to some of the
extended ranges and other features that are now available using the default SCPI command set. In
addition, some Series 2000 or Keysight 34401 code will work differently in the DMM6500 than it did in
the earlier instrument. For information about the differences between the DMM6500 and the Series
2000, refer to Keithley Instruments document #0771466, DMM6500 in a Model 2000 application. For
information about the differences between the DMM6500 and the Keysight 34401, refer to Keithley
Instruments document #0771467, DMM6500 in a Keysight Model 34401 application.
To set the command set from the front panel:
1. Press the MENU key.
2. Under System, select Settings.
3. Select Command Set.
You are prompted to reboot.
DMM6500-900-01Rev. A / April 2018
®
®
Windows
, you can check the file system by checking the properties of the USB flash
Section 10: Troubleshooting FAQs
10-3