Commands by Subsystem

:POD Commands

Control all oscilloscope functions associated with groups of digital
channels. See
Table 71 :POD<n> Commands Summary
Command
:POD<n>:DISPlay {{0 |
OFF} | {1 | ON}} (see
page
393)
:POD<n>:SIZE <value>
(see
page
394)
:POD<n>:THReshold
<type>[suffix] (see
page
395)
Introduction to
<n> ::= {1 | 2}
:POD<n>
The POD subsystem commands control the viewing and threshold of
Commands
groups of digital channels.
POD1 ::= D0- D7
POD2 ::= D8- D15
These commands are only valid for the MSO models.
N O T E
Reporting the Setup
Use :POD1? or :POD2? to query setup information for the POD subsystem.
Return Format
The following is a sample response from the :POD1? query. In this case,
the query was issued following a *RST command.
:POD1:DISP 0;THR +1.40E+00
392
"Introduction to :POD<n> Commands"
Query
:POD<n>:DISPlay? (see
page
393)
:POD<n>:SIZE? (see
page
394)
:POD<n>:THReshold?
(see
page
395)
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
on page 392.
Options and Query Returns
{0 | 1}
<n> ::= 1-2 in NR1 format
<value> ::= {SMALl | MEDium |
LARGe}
<n> ::= 1-2 in NR1 format
<type> ::= {CMOS | ECL | TTL |
<user defined value>}
<user defined value> ::= value in
NR3 format
[suffix] ::= {V | mV | uV }