Autoscale - Agilent Technologies InfiniiVision 7000A Series Programmer's Manual

Table of Contents

Advertisement

5
Commands by Subsystem

:AUToscale

Command Syntax
:AUToscale
:AUToscale [<source>[,..,<source>]]
<source> ::= CHANnel<n> for the DSO models
<source> ::= {DIGital0,..,DIGital15 | POD1 | POD2 | CHANnel<n>} for the
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The <source> parameter may be repeated up to 5 times.
The :AUToscale command evaluates all input signals and sets the correct
conditions to display the signals. This is the same as pressing the
Autoscale key on the front panel.
If one or more sources are specified, those specified sources will be
enabled and all others blanked. The autoscale channels mode (see
":AUToscale:CHANnels"
the autoscale is performed.
When the :AUToscale command is sent, the following conditions are
affected and actions are taken:
• Thresholds.
• Channels with activity around the trigger point are turned on, others
• Channels are reordered on screen; analog channel 1 first, followed by
• Delay is set to 0 seconds.
• Time/Div.
The :AUToscale command does not affect the following conditions:
• Label names.
• Trigger conditioning.
The :AUToscale command turns off the following items:
• Cursors.
• Measurements.
• Trace memories.
• Zoomed (delayed) time base mode.
For further information on :AUToscale, see the User's Guide.
144
(see
page
786)
MSO models
are turned off.
the remaining analog channels, then the digital channels 0- 15.
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
on page 147) is set to DISPlayed channels. Then,

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents