:MEASure:SOURce

Command Syntax
:MEASure:SOURce <source1>[,<source2>]
<source1>,<source2> ::= {<digital channels> | CHANnel<n> | FUNCtion
<digital channels> ::= DIGital0,..,DIGital15 for the MSO models
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MEASure:SOURce command sets the default sources for
measurements. The specified sources are used as the sources for the
MEASure subsystem commands if the sources are not explicitly set with
the command.
If a source is specified for any measurement, the current source is
changed to this new value.
If :MARKer:MODE is set to OFF or MANual, setting :MEASure:SOURce to
CHANnel<n>, FUNCtion, or MATH will also set :MARKer:X1Y1source to
source1 and :MARKer:X2Y2source to source2.
EXTernal is only a valid source for the counter measurement (and
<source1>).
Query Syntax
:MEASure:SOURce?
The :MEASure:SOURce? query returns the current source selections. If
source2 is not specified, the query returns "NONE" for source2. If all
channels are off, the query returns "NONE,NONE". Source2 only applies to
:MEASure:DELay and :MEASure:PHASe measurements.
MATH is an alias for FUNCtion. The query will return FUNC if the source is FUNCtion or
N O T E
MATH.
Return Format
<source1>,<source2><NL>
<source1>,<source2> ::= {<digital channels> | CHAN<n> | FUNC | EXT
See Also:
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
(see
page
786)
"Introduction to :MEASure Commands"
":MARKer:MODE"
on page 297
":MARKer:X1Y1source"
":MARKer:X2Y2source"
":MEASure:DELay"
on page 319
":MEASure:PHASe"
on page 328
| MATH | EXTernal}
| NONE}
on page 312
on page 299
on page 301