:TRIGger:CAN:PATTern:ID:MODE

Command Syntax
:TRIGger:CAN:PATTern:ID:MODE <value>
<value> ::= {STANdard | EXTended}
The :TRIGger:CAN:PATTern:ID:MODE command sets the CAN identifier
mode. STANdard selects the standard 11- bit identifier. EXTended selects
the extended 29- bit identifier. The CAN identifier is set by the
:TRIGger:CAN:PATTern:ID command.
This command is only valid when the automotive CAN and LIN serial decode option (Option
N O T E
AMS) has been licensed.
Query Syntax
:TRIGger:CAN:PATTern:ID:MODE?
The :TRIGger:CAN:PATTern:ID:MODE? query returns the current setting of
the CAN identifier mode.
Return Format
<value><NL>
<value> ::= {STAN | EXT}
Errors
See Also
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
(see
page
786)
"- 241, Hardware missing"
"Introduction to :TRIGger Commands"
":TRIGger:MODE"
on page 473
":TRIGger:CAN:PATTern:DATA"
":TRIGger:CAN:PATTern:DATA:LENGth"
":TRIGger:CAN:PATTern:ID"
on page 745
on page 467
on page 480
on page 481
on page 482