Marker:x2Y2Source - Agilent Technologies InfiniiVision 7000A Series Programmer's Manual

Table of Contents

Advertisement

:MARKer:X2Y2source

Command Syntax
:MARKer:X2Y2source <source>
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MARKer:X2Y2source command sets the source for the cursors. The
channel you specify must be enabled for cursors to be displayed. If the
channel or function is not on, an error message is issued.
If the marker mode is not currently WAVeform (see
page 297):
• Sending a :MARKer:X2Y2source command will put the cursors in the
• Setting the source for one pair of markers (for example, X2Y2) sets the
If the marker mode is currently WAVeform, the X2Y2 source can be set
separate from the X1Y1 source.
If :MARKer:MODE is set to OFF or MANual, setting :MEASure:SOURce to
CHANnel<n>, FUNCtion, or MATH will also set :MARKer:X1Y1source and
:MARKer:X2Y2source to this value.
MATH is an alias for FUNCtion. The query will return FUNC if the source is FUNCtion or
N O T E
MATH.
Query Syntax
:MARKer:X2Y2source?
The :MARKer:X2Y2source? query returns the current source for the
cursors. If all channels are off or if :MARKer:MODE is set to OFF, the
query returns NONE.
Return Format
<source><NL>
<source> ::= {CHAN<n> | FUNC | NONE}
See Also
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
(see
page
786)
MANual mode.
source for the other (for example, X1Y1).
"Introduction to :MARKer Commands"
":MARKer:MODE"
on page 297
":MARKer:X1Y1source"
":MEASure:SOURce"
on page 296
on page 299
on page 337
Commands by Subsystem
":MARKer:MODE"
on
5
301

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents