Other Ways of Sending Commands

Standard Commands for Programmable Instrumentation (SCPI) can be sent
via a Telnet socket or through the Browser Web Control.

Telnet Sockets

The following information is provided for programmers who wish to
control the oscilloscope with SCPI commands in a Telnet session.
To connect to the oscilloscope via a telnet socket, issue the following
command:
telnet <hostname> 5024
where <hostname> is the hostname of the oscilloscope. This will give you a
command line with prompt.
For a command line without a prompt, use port 5025. For example:
telnet <hostname> 5025

Sending SCPI Commands Using Browser Web Control

To send SCPI commands using the Browser Web Control feature, establish
a connection to the oscilloscope via LAN as described in the 7000 Series
Oscilloscopes User's Guide. When you make the connection to the
oscilloscope via LAN and the instrument's welcome page is displayed,
select the Browser Web Control tab, then select the Remote
Programming link.