Mtest:amask:source - Agilent Technologies InfiniiVision 7000A Series Programmer's Manual

Table of Contents

Advertisement

5
Commands by Subsystem

:MTESt:AMASk:SOURce

Command Syntax
:MTESt:AMASk:SOURce <source>
<source> ::= CHANnel<n>
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MTESt:AMASk:SOURce command selects the source for the
interpretation of the :MTESt:AMASk:XDELta and :MTESt:AMASk:YDELta
parameters when :MTESt:AMASk:UNITs is set to CURRent.
When UNITs are CURRent, the XDELta and YDELta parameters are
defined in terms of the channel units, as set by the :CHANnel<n>:UNITs
command, of the selected source.
Suppose that UNITs are CURRent and that you set SOURce to CHANNEL1,
which is using units of volts. Then you can define AMASk:XDELta in
terms of volts and AMASk:YDELta in terms of seconds.
This command is the same as the :MTESt:SOURce command.
Query Syntax
:MTESt:AMASk:SOURce?
The :MTESt:AMASk:SOURce? query returns the currently set source.
Return Format
<source> ::= CHAN<n>
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
See Also
Example Code
364
(see
page
786)
"Introduction to :MTESt Commands"
":MTESt:AMASk:XDELta"
":MTESt:AMASk:YDELta"
":MTESt:AMASk:UNITs"
":MTESt:SOURce"
on page 390
"Example Code"
on page 360
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
on page 360
on page 366
on page 367
on page 365

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents