:TRIGger:SEQuence:RESet

Command Syntax
:TRIGger:SEQuence:RESet <value>
<value> ::= {NONE | PATTern1,ENTered | PATTern1,EXITed | EDGE1
Values used in find and trigger stages are not available.
not available if EDGE2,COUNt is used in trigger stage.
The :TRIGger:SEQuence:RESet command specifies the reset stage of a
sequence trigger. In multi- level trigger specifications, you may find a
pattern, then search for another in sequence, but reset the entire search to
the beginning if another condition occurs. This command accepts three
program data parameters; you can use NONE to fill out the parameter list
(for example, "EDGE1,NONE,NONE").
PATTern1 and PATTern2 are specified with the
:TRIGger:SEQuence:PATTern command. EDGE1 and EDGE2 are specified
with the :TRIGger:SEQuence:EDGE command. TIMer is specified with the
:TRIGger:SEQuence:TIMer command.
Query Syntax
:TRIGger:SEQuence:RESet?
The :TRIGger:SEQuence:RESet? query returns the reset stage specification
for a sequence trigger. NONE is returned for unused parameters.
Return Format
<reset_value><NL>
<reset_value> ::= {NONE,NONE,NONE | PATT1,ENT,NONE | PATT1,EXIT,NONE
See Also
578
(see
page
786)
| PATTern1,AND,EDGE1 | PATTern2,ENTered | PATTern2,EXITed
| EDGE2 | TIMer}
| EDGE1,NONE,NONE | PATT1,AND,EDGE1 | PATT2,ENT,NONE
| PATT2,EXIT,NONE | EDGE2,NONE,NONE | TIM,NONE,NONE}
"Introduction to :TRIGger Commands"
":TRIGger:SEQuence:PATTern"
":TRIGger:SEQuence:EDGE"
":TRIGger:SEQuence:TIMer"
":TRIGger:SEQuence:FIND"
":TRIGger:SEQuence:TRIGger"
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide