Hweregister[:Event] (Hardware Event Event Register) - Agilent Technologies InfiniiVision 7000A Series Programmer's Manual

Table of Contents

Advertisement

5
Commands by Subsystem

:HWERegister[:EVENt] (Hardware Event Event Register)

Query Syntax
:HWERegister[:EVENt]?
The :HWERegister[:EVENt]? query returns the integer value contained in
the Hardware Event Event Register.
PLL
Locked
12
PLL
Locked
15
14
13
12
+
OR
To HWE bit in
Operation Status
Condition Register
Table 48 Hardware Event Event Register
Bit
15-13
12
11-1
0
Return Format
<value><NL>
<value> ::= integer in NR1 format.
See Also
156
(see
page
786)
11
10
9
8
7
Name
Description
---
---
PLL
PLL Locked
Locked
---
---
Bat On
Battery On
"Introduction to Root (:) Commands"
":CHANnel<n>:PROTection"
":EXTernal:PROTection"
":OPEE (Operation Status Enable Register)"
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
6
5
4
3
2
When Set (1 = High = True), Indicates:
(Not used.)
This bit is for internal use and is not intended
for general use.
(Not used.)
The battery is on.
on page 141
on page 232
on page 261
Bat
:HWERegister:CONDition?
Hardware Event Condition Register
ON
0
Bat
:HWERegister[:EVENt]?
Hardware Event Event Register
ON
1
0
:HWEenable
:HWEenable?
Hardware Event Enable (Mask) Register
on page 163