:CALibrate Commands

Utility commands for viewing calibration status and for starting the user
calibration procedure. See
page 207.
Table 58 :CALibrate Commands Summary
Command
n/a
:CALibrate:LABel
<string> (see
page
210)
:CALibrate:OUTPut
<signal> (see
page
211)
:CALibrate:STARt (see
page
212)
n/a
n/a
n/a
n/a
Introduction to
The CALibrate subsystem provides utility commands for:
:CALibrate
• Determining the state of the calibration factor protection switch
Commands
• Saving and querying the calibration label string.
• Reporting the calibration time and date.
• Reporting changes in the temperature since the last calibration.
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
Query
:CALibrate:DATE? (see
page
209)
:CALibrate:LABel?
(see
page
210)
:CALibrate:OUTPut?
(see
page
211)
n/a
:CALibrate:STATus?
(see
page
213)
:CALibrate:SWITch?
(see
page
214)
:CALibrate:TEMPeratur
e? (see
page
:CALibrate:TIME? (see
page
216)
(CAL PROTECT).
"Introduction to :CALibrate Commands"
Options and Query Returns
<return value> ::=
<day>,<month>,<year>; all in NR1
format
<string> ::= quoted ASCII string
up to 32 characters
<signal> ::= {TRIGgers | SOURce |
DSOurce | MASK}
n/a
<return value> ::=
ALL,<status_code>,<status_string
>
<status_code> ::= an integer
status code
<status_string> ::= an ASCII
status string
{PROTected | UNPRotected}
<return value> ::= degrees C
215)
delta since last cal in NR3
format
<return value> ::=
<hours>,<minutes>,<seconds>; all
in NR1 format