7
Obsolete and Discontinued Commands
:CHANnel<n>:INPut
(see
page
786)
Command Syntax
:CHANnel<n>:INPut <impedance>
<impedance> ::= {ONEMeg | FIFTy}
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :CHANnel<n>:INPut command selects the input impedance setting for
the specified channel. The legal values for this command are ONEMeg (1
Ω
Ω
M
) and FIFTy (50
).
The :CHANnel<n>:INPut command is an obsolete command provided for compatibility to
N O T E
previous oscilloscopes. Use the :CHANnel<n>:IMPedance command (see
page
223)
instead.
Query Syntax
:CHANnel<n>:INPut?
The :CHANnel<n>:INPut? query returns the current input impedance
setting for the specified channel.
Return Format
<impedance value><NL>
<impedance value> ::= {ONEM | FIFT}
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide