Operegister[:Event] (Operation Status Event Register) - Agilent Technologies InfiniiVision 7000A Series Programmer's Manual

Table of Contents

Advertisement

:OPERegister[:EVENt] (Operation Status Event Register)

Query Syntax
:OPERegister[:EVENt]?
The :OPERegister[:EVENt]? query returns the integer value contained in
the Operation Status Event Register.
From Hardware
Event Registers
From Overload
Event Registers
HWE
OVLR
12
HWE
OVLR
15
14
13
12
+
OR
To OPER bit in
Status Byte
Register
Table 53 Operation Status Event Register
Bit
15-13
12
11
10
9
8-6
5
4
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
(see
page
786)
From Mask Test
Event Registers
MTE
9
11
MTE
11
10
9
8
7
6
Name
Description
---
---
HWE
Hardware Event
OVLR
Overload
---
---
MTE
Mask Test Event
---
---
Wait
Wait Trig
Trig
---
---
Arm
AER?
Reg
Run bit set if oscilloscope not stopped
Wait
Run
Trig
5
3
Wait
Run
Trig
5
4
3
2
1
When Set (1 = High = True), Indicates:
(Not used.)
A hardware event has occurred.
A 50Ω input overload has occurred.
(Not used.)
A mask test event has occurred.
(Not used.)
The trigger is armed (set by the Trigger Armed
Event Register (TER)).
(Not used.)
Commands by Subsystem
:OPEReration:CONDition?
Operation Status Condition Register
:OPERation[:EVENt]?
Operation Status Event Register
0
:OPEE
:OPEE?
Operation Status Enable
(Mask) Register
5