:CHANnel:ACTivity

Command Syntax
:CHANnel:ACTivity
The :CHANnel:ACTivity command clears the cumulative edge variables for
the next activity query.
The :CHANnel:ACTivity command is an obsolete command provided for compatibility to
N O T E
previous oscilloscopes. Use the :ACTivity command (see
Query Syntax
:CHANnel:ACTivity?
The :CHANnel:ACTivity? query returns the active edges since the last clear,
and returns the current logic levels.
Return Format
<edges>,<levels><NL>
<edges> ::= presence of edges (32-bit integer in NR1 format).
<levels> ::= logical highs or lows (32-bit integer in NR1 format).
A bit equal to zero indicates that no edges were detected at the specified threshold since
N O T E
the last clear on that channel. Edges may have occurred that were not detected because of
the threshold setting.
A bit equal to one indicates that edges have been detected at the specified
threshold since the last clear on that channel.
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
(see
page
786)
Obsolete and Discontinued Commands
page
142) instead.