:TIMebase:MODE

Command Syntax
:TIMebase:MODE <value>
<value> ::= {MAIN | WINDow | XY | ROLL}
The :TIMebase:MODE command sets the current time base. There are four
time base modes:
• MAIN — The normal time base mode is the main time base. It is the
• WINDow — In the WINDow (zoomed or delayed) time base mode,
• XY — In the XY mode, the :TIMebase:RANGe, :TIMebase:POSition, and
• ROLL — In the ROLL mode, data moves continuously across the display
If a :DIGitize command is executed when the :TIMebase:MODE is not MAIN, the
N O T E
:TIMebase:MODE is set to MAIN.
Query Syntax
:TIMebase:MODE?
The :TIMebase:MODE query returns the current time base mode.
Return Format
<value><NL>
<value> ::= {MAIN | WIND | XY | ROLL}
See Also
Example Code
Example program from the start:
page 812
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
(see
page
786)
default time base mode after the *RST (Reset) command.
measurements are made in the zoomed time base if possible; otherwise,
the measurements are made in the main time base.
:TIMebase:REFerence commands are not available. No measurements are
available in this mode.
from left to right. The oscilloscope runs continuously and is untriggered.
The :TIMebase:REFerence selection changes to RIGHt.
"Introduction to :TIMebase Commands"
"*RST (Reset)"
on page 128
":TIMebase:RANGe"
on page 459
":TIMebase:POSition"
":TIMebase:REFerence"
' TIMEBASE_MODE - (not executed in this example)
' Set the time base mode to MAIN, DELAYED, XY, or ROLL.
' Set time base mode to main.
myScope.WriteString ":TIMEBASE:MODE MAIN"