:TRIGger Commands

Control the trigger modes and parameters for each trigger type. See:
Introduction to
The commands in the TRIGger subsystem define the conditions for an
:TRIGger
internal trigger. Many of these commands are valid in multiple trigger
Commands
modes.
The default trigger mode is :EDGE.
The trigger subsystem controls the trigger sweep mode and the trigger
specification. The trigger sweep (see
be AUTO or NORMal.
• NORMal mode — displays a waveform only if a trigger signal is present
• AUTO trigger mode — generates an artificial trigger event if the trigger
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
"Introduction to :TRIGger Commands"
"General :TRIGger Commands"
":TRIGger:CAN Commands"
":TRIGger:DURation Commands"
":TRIGger:EBURst Commands"
":TRIGger[:EDGE] Commands"
":TRIGger:FLEXray Commands"
":TRIGger:GLITch Commands"
":TRIGger:I2S Commands"
":TRIGger:IIC Commands"
":TRIGger:LIN Commands"
":TRIGger:M1553 Commands"
":TRIGger:SEQuence Commands"
":TRIGger:SPI Commands"
":TRIGger:TV Commands"
":TRIGger:USB Commands"
":TRIGger:UART Commands"
and the trigger conditions are met. Otherwise the oscilloscope does not
trigger and the display is not updated. This mode is useful for
low- repetitive- rate signals.
specification is not satisfied within a preset time, acquires
unsynchronized data and displays it.
AUTO mode is useful for signals other than low- repetitive- rate signals.
You must use this mode to display a DC signal because there are no
edges on which to trigger.
on page 467
on page 470
on page 478
on page 489
on page 495
on page 499
on page 505
on page 517 (Pulse Width trigger)
on page 526
on page 544
on page 553
on page 566
on page 573
on page 581
on page 590
on page 611
on page 596
":TRIGger:SWEep"
Commands by Subsystem
on page 477) can
5
467