:TRIGger:UART:BAUDrate

Command Syntax
:TRIGger:UART:BAUDrate <baudrate>
<baudrate> ::= integer from 1200 to 3000000 in 100 b/s increments
The :TRIGger:UART:BAUDrate command selects the bit rate (in bps) for the
serial decoder and/or trigger when in UART mode. The baud rate can be
set from 1200 b/s to 3 Mb/s in 100 b/s increments. If you enter a baud
rate that is not divisible by 100 b/s, the baud rate is set to the nearest
baud rate divisible by 100 b/s.
If the baud rate you select does not match the system baud rate, false
triggers may occur.
Query Syntax
:TRIGger:UART:BAUDrate?
The :TRIGger:UART:BAUDrate? query returns the current UART baud rate
setting.
Return Format
<baudrate><NL>
<baudrate> ::= integer from 1200 to 3000000 in 100 b/s increments
See Also
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
(see
page
786)
"Introduction to :TRIGger Commands"
":TRIGger:MODE"
on page 473
":TRIGger:UART:TYPE"
on page 467
on page 609
Commands by Subsystem