5
Commands by Subsystem
CAN Id specification is set by the :TRIGger:CAN:PATTern:ID
and:TRIGger:CAN:PATTern:ID:MODE commands.
CAN Data specification is set by the :TRIGger:CAN:PATTern:DATA
command.
CAN Data Length Code is set by the :TRIGger:CAN:PATTern:DATA:LENGth
command.
SOF is the only valid selection for analog oscilloscopes. If the automotive CAN and LIN
N O T E
serial decode option (Option AMS) has not been licensed, SOF is the only valid selection.
Query Syntax
:TRIGger:CAN:TRIGger?
The :TRIGger:CAN:TRIGger? query returns the current CAN trigger on
condition.
Return Format
<condition><NL>
<condition> ::= {SOF | DATA | ERR | IDD | IDE | IDR | ALL | OVER | ACK}
Errors
•
See Also
•
•
•
•
•
•
•
•
488
"- 241, Hardware missing"
"Introduction to :TRIGger Commands"
":TRIGger:MODE"
on page 473
":TRIGger:CAN:PATTern:DATA"
":TRIGger:CAN:PATTern:DATA:LENGth"
":TRIGger:CAN:PATTern:ID"
":TRIGger:CAN:PATTern:ID:MODE"
":TRIGger:CAN:SIGNal:DEFinition"
":TRIGger:CAN:SOURce"
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
on page 745
on page 467
on page 480
on page 481
on page 482
on page 483
on page 739
on page 486