5
Commands by Subsystem

:MEASure Commands

Select automatic measurements to be made and control time markers. See
"Introduction to :MEASure Commands"
Table 69 :MEASure Commands Summary
Command
:MEASure:CLEar (see
page
314)
:MEASure:COUNter
[<source>] (see
page
315)
:MEASure:DEFine
DELay, <delay spec>
(see
page
316)
:MEASure:DEFine
THResholds,
<threshold spec> (see
page
316)
:MEASure:DELay
[<source1>]
[,<source2>] (see
page
319)
:MEASure:DUTYcycle
[<source>] (see
page
321)
306
Query
n/a
:MEASure:COUNter?
[<source>] (see
page
315)
:MEASure:DEFine?
DELay (see
page
317)
:MEASure:DEFine?
THResholds (see
page
317)
:MEASure:DELay?
[<source1>]
[,<source2>] (see
page
319)
:MEASure:DUTYcycle?
[<source>] (see
page
321)
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
on page 312.
Options and Query Returns
n/a
<source> ::= {CHANnel<n> |
EXTernal} for DSO models
<source> ::= {CHANnel<n> |
DIGital0,..,DIGital15 |
EXTernal} for MSO models
<n> ::= 1-2 or 1-4 in NR1 format
<return_value> ::= counter
frequency in Hertz in NR3 format
<delay spec> ::=
<edge_spec1>,<edge_spec2>
edge_spec1 ::=
[<slope>]<occurrence>
edge_spec2 ::=
[<slope>]<occurrence>
<slope> ::= {+ | -}
<occurrence> ::= integer
<threshold spec> ::= {STANdard} |
{<threshold mode>,<upper>,
<middle>,<lower>}
<threshold mode> ::= {PERCent |
ABSolute}
<source1,2> ::= {CHANnel<n> |
FUNCtion | MATH}
<n> ::= 1-2 or 1-4 in NR1 format
<return_value> ::=
floating-point number delay time
in seconds in NR3 format
<source> ::= {CHANnel<n> |
FUNCtion | MATH} for DSO models
<source> ::= {CHANnel<n> |
DIGital0,..,DIGital15 | FUNCtion
| MATH} for MSO models
<n> ::= 1-2 or 1-4 in NR1 format
<return_value> ::= ratio of
positive pulse width to period in
NR3 format