:MEASure:VSTArt

Command Syntax
:MEASure:VSTArt <vstart_argument>
<vstart_argument> ::= value for vertical marker 1
The :MEASure:VSTArt command moves the vertical marker (Y1 cursor) to
the specified value corresponding to the selected source. The source can
be selected by the MARKer:X1Y1source command.
The short form of this command, VSTA, does not follow the defined Long Form to Short
N O T E
Form Truncation Rules (see
both VSTArt and VSTOp, so sending VST for the VSTArt command produces an error.
The :MEASure:VSTArt command is an obsolete command provided for compatibility to
N O T E
previous oscilloscopes. Use the :MARKer:Y1Position command (see
Query Syntax
:MEASure:VSTArt?
The :MEASure:VSTArt? query returns the current value of the Y1 cursor.
Return Format
<value><NL>
<value> ::= voltage at voltage marker 1 in NR3 format
See Also
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
(see
page
786)
"Introduction to :MARKer Commands"
"Introduction to :MEASure Commands"
":MARKer:Y1Position"
":MARKer:Y2Position"
":MARKer:YDELta"
on page 305
":MARKer:X1Y1source"
":MEASure:SOURce"
":MEASure:TDELta"
on page 715
":MEASure:TSTArt"
on page 719