5
Commands by Subsystem

:TRIGger:IIC[:SOURce]:CLOCk

Command Syntax
:TRIGger:IIC:[SOURce:]CLOCk <source>
<source> ::= {CHANnel<n> | EXTernal} for the DSO models
<source> ::= {CHANnel<n> | DIGital0,..,DIGital15} for the MSO models
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :TRIGger:IIC:[SOURce:]CLOCk command sets the source for the IIC
serial clock (SCL).
Query Syntax
:TRIGger:IIC:[SOURce:]CLOCk?
The :TRIGger:IIC:[SOURce:]CLOCk? query returns the current source for
the IIC serial clock.
Return Format
<source><NL>
See Also
548
(see
page
786)
"Introduction to :TRIGger Commands"
":TRIGger:IIC[:SOURce]:DATA"
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
on page 467