:DIGital<n>:THReshold
Command Syntax
:DIGital<n>:THReshold <value>
<value> ::= {CMOS | ECL | TTL | <user defined value>[<suffix>]}
<user defined value> ::= -8.00 to +8.00 in NR3 format
<suffix> ::= {V | mV | uV}
<n> ::= An integer, 0, 1,..,15, is attached as a suffix to the command
and defines the logic channel that is affected by the command.
• TTL = 1.4V
• CMOS = 2.5V
• ECL = - 1.3V
The :DIGital<n>:THReshold command sets the logic threshold value for all
channels grouped with the specified channel (D0- D7, D8- D15). The
threshold is used for triggering purposes and for displaying the digital
data as high (above the threshold) or low (below the threshold).
This command is only valid for the MSO models.
N O T E
Query Syntax
:DIGital<n>:THReshold?
The :DIGital<n>:THReshold? query returns the threshold value for the
specified channel.
Return Format
<value><NL>
<value> ::= threshold value in NR3 format
See Also
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
(see
page
786)
"Introduction to :DIGital<n> Commands"
":POD<n>:THReshold"
":TRIGger[:EDGE]:LEVel"
on page 237
on page 395
on page 501