:TRIGger:FLEXray:EVENt:TYPE

Command Syntax
:TRIGger:FLEXray:EVENt:TYPE <event>
<event> ::= {WAKeup | TSS | {FES | DTS} | BSS}
Selects the FlexRay event to trigger on. The event setting is only valid
when the FlexRay trigger mode is set to EVENt.
• WAKeup — triggers on Wake- Up event.
• TSS — triggers on Transmission Start Sequence event.
• FES — triggers on Frame End Sequence event.
• DTS — triggers on Dynamic Trailing Sequence event.
• BSS — triggers on Byte Start Sequence event.
FES and DTS are equivalent.
N O T E
This command is only valid when the FLEXray triggering and serial decode option (Option
N O T E
FLX) has been licensed.
Query Syntax
:TRIGger:FLEXray:EVENt:TYPE?
The :TRIGger:FLEXray:EVENt:TYPE? query returns the currently selected
FLEXray event.
Return Format
<event><NL>
<event> ::= {WAK | TSS | {FES | DTS} | BSS}
See Also
510
(see
page
786)
"Introduction to :TRIGger Commands"
":TRIGger:FLEXray:TRIGger"
":TRIGger:FLEXray:AUTosetup"
":TRIGger:FLEXray:SOURce"