Table 74 :SBUS Commands Summary (continued)
Command
:SBUS:SPI:BITorder
<order> (see
page
437)
:SBUS:SPI:WIDTh
<word_width> (see
page
438)
:SBUS:UART:BASE
<base> (see
page
n/a
:SBUS:UART:COUNt:RESe
t (see
page
441)
n/a
n/a
:SBUS:UART:FRAMing
<value> (see
page
444)
Introduction to
The :SBUS subsystem commands control the serial decode bus viewing,
:SBUS
mode, and other options.
Commands
These commands are only valid on 4 (analog) channel oscilloscope models when a serial
N O T E
decode option has been licensed.
Reporting the Setup
Use :SBUS? to query setup information for the :SBUS subsystem.
Return Format
The following is a sample response from the :SBUS? query. In this case,
the query was issued following a *RST command.
:SBUS:DISP 0;MODE IIC
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
Query
:SBUS:SPI:BITorder?
(see
page
437)
:SBUS:SPI:WIDTh? (see
page
438)
:SBUS:UART:BASE? (see
439)
page
439)
:SBUS:UART:COUNt:ERRo
r? (see
page
n/a
:SBUS:UART:COUNt:RXFR
ames? (see
page
:SBUS:UART:COUNt:TXFR
ames? (see
page
:SBUS:UART:FRAMing?
(see
page
444)
Options and Query Returns
<order> ::= {LSBFirst | MSBFirst}
<word_width> ::= integer 4-16 in
NR1 format
<base> ::= {ASCii | BINary | HEX}
<frame_count> ::= integer in NR1
440)
format
n/a
<frame_count> ::= integer in NR1
442)
format
<frame_count> ::= integer in NR1
443)
format
<value> ::= {OFF | <decimal> |
<nondecimal>}
<decimal> ::= 8-bit integer from
0-255 (0x00-0xff)
<nondecimal> ::= #Hnn where n ::=
{0,..,9 | A,..,F} for hexadecimal
<nondecimal> ::= #Bnn...n where n
::= {0 | 1} for binary