Lrn (Learn Device Setup) - Agilent Technologies InfiniiVision 7000A Series Programmer's Manual

Table of Contents

Advertisement

*LRN (Learn Device Setup)

Query Syntax
*LRN?
The *LRN? query result contains the current state of the instrument. This
query is similar to the :SYSTem:SETup? (see
it contains ":SYST:SET " before the binary block data. The query result is a
valid command that can be used to restore instrument settings at a later
time.
Return Format
<learn_string><NL>
<learn_string> ::= :SYST:SET <setup_data>
<setup_data> ::= binary block data in IEEE 488.2 # format
<learn string> specifies the current instrument setup. The block size is
subject to change with different firmware revisions.
The *LRN? query return format has changed from previous Agilent oscilloscopes to match
N O T E
the IEEE 488.2 specification which says that the query result must contain ":SYST:SET "
before the binary block data.
See Also
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
(see
page
786)
"Introduction to Common (*) Commands"
"*RCL (Recall)"
on page 127
"*SAV (Save)"
on page 131
":SYSTem:SETup"
on page 452
Commands by Subsystem