Trigger[:Edge]:Coupling - Agilent Technologies InfiniiVision 7000A Series Programmer's Manual

Table of Contents

Advertisement

5
Commands by Subsystem

:TRIGger[:EDGE]:COUPling

Command Syntax
:TRIGger[:EDGE]:COUPling <coupling>
<coupling> ::= {AC | DC | LFReject}
The :TRIGger[:EDGE]:COUPling command sets the input coupling for the
selected trigger sources. The coupling can be set to AC, DC, or LFReject.
• AC coupling places a high- pass filter (10 Hz for analog channels, and
• LFReject coupling places a 50 KHz high- pass filter in the trigger path.
• DC coupling allows dc and ac signals into the trigger path.
The :TRIGger[:EDGE]:COUPling and the :TRIGger[:EDGE]:REJect selections are coupled.
N O T E
Changing the setting of the :TRIGger[:EDGE]:REJect can change the COUPling setting.
Query Syntax
:TRIGger[:EDGE]:COUPling?
The :TRIGger[:EDGE]:COUPling? query returns the current coupling
selection.
Return Format
<coupling><NL>
<coupling> ::= {AC | DC | LFR}
See Also
500
(see
page
786)
3.5 Hz for all External trigger inputs) in the trigger path, removing dc
offset voltage from the trigger waveform. Use AC coupling to get a
stable edge trigger when your waveform has a large dc offset.
"Introduction to :TRIGger Commands"
":TRIGger:MODE"
on page 473
":TRIGger[:EDGE]:REJect"
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
on page 467
on page 502