:ACQuire:RSIGnal

Command Syntax
:ACQuire:RSIGnal <ref_signal_mode>
<ref_signal_mode> ::= {OFF | OUT | IN}
The :ACQuire:RSIGnal command selects the 10 MHz reference signal mode.
• The OFF mode disables the oscilloscope's 10 MHz REF BNC connector.
• The OUT mode is used to synchronize the timebase of two or more
• The IN mode is used to supply a sample clock to the oscilloscope. A
Do not apply more than ±15 V at the 10 MHz REF BNC connector on the rear panel,
C A U T I O N
or damage to the instrument may occur.
Query Syntax
:ACQuire:RSIGnal?
The :ACQuire:RSIGnal? query returns the current 10 MHz reference signal
mode.
Return Format
<ref_signal_mode><NL>
<ref_signal_mode> ::= {OFF | OUT | IN}
See Also
• The Agilent InfiniiVision 7000 Series Oscilloscope User's Guide for
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
(see
page
786)
instruments.
10 MHz square or sine wave signal is input to the BNC connector
labeled 10 MHz REF. The amplitude must be between 180 mV and 1 V,
with an offset of between 0 V and 2 V.
":TIMebase:REFClock"
information on using the 10 MHz reference clock.
on page 460
Commands by Subsystem