Trigger:flexray:error:type - Agilent Technologies InfiniiVision 7000A Series Programmer's Manual

Table of Contents

Advertisement

:TRIGger:FLEXray:ERRor:TYPE

Command Syntax
:TRIGger:FLEXray:ERRor:TYPE <error_type>
<error_type> ::= {ALL | HCRC | FCRC}
Selects the FlexRay error type to trigger on. The error type setting is only
valid when the FlexRay trigger mode is set to ERRor.
• ALL — triggers on ALL errors.
• HCRC — triggers on only Header CRC errors.
• FCRC — triggers on only Frame CRC errors.
This command is only valid when the FLEXray triggering and serial decode option (Option
N O T E
FLX) has been licensed.
Query Syntax
:TRIGger:FLEXray:ERRor:TYPE?
The :TRIGger:FLEXray:ERRor:TYPE? query returns the currently selected
FLEXray error type.
Return Format
<error_type><NL>
<error_type> ::= {ALL | HCRC | FCRC}
See Also
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
(see
page
786)
"Introduction to :TRIGger Commands"
":TRIGger:FLEXray:TRIGger"
Commands by Subsystem
on page 467
on page 516
5
509