System:precision - Agilent Technologies InfiniiVision 7000A Series Programmer's Manual

Table of Contents

Advertisement

5
Commands by Subsystem

:SYSTem:PRECision

Command Syntax
:SYSTem:PRECision <value>
<value> ::= {{1 | ON} | {0 | OFF}}
The :SYSTem:PRECision command turns the oscilloscope's precision
analysis setting on or off.
• OFF (0) — provides the maximum oscilloscope waveform update rate by
• ON (1) — at the expense of oscilloscope waveform update rate, this
The precision analysis setting is OFF after a *RST command.
Precision analysis is not available when:
• Realtime sampling mode is off.
• Averaging or High Resolution acquisition modes are selected.
• XY or Roll time modes are selected.
Query Syntax
:SYSTem:PRECision?
The :SYSTem:PRECision? query returns the current precision analysis
setting.
Return Format
<value><NL>
<value> ::= {1 | 0}
See Also
450
(see
page
786)
performing measurements and math functions on a 1000- point
measurement record.
setting allows measurements and math functions to be performed on a
precision analysis record (see
"Introduction to :SYSTem Commands"
":WAVeform:POINts:MODE"
"*RST (Reset)"
on page 128
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
":WAVeform:POINts:MODE"
on page 445
on page 631
on page 631).