5
Commands by Subsystem

:MARKer:Y2Position

Command Syntax
:MARKer:Y2Position <position> [suffix]
<position> ::= Y2 cursor position in NR3 format
<suffix> ::= {mV | V | dB}
If the :MARKer:MODE is not currently set to WAVeform (see
":MARKer:MODE"
• Sets :MARKer:MODE to MANual.
• Sets the Y2 cursor position to the specified value.
When the :MARKer:MODE is set to WAVeform, Y positions cannot be set.
Query Syntax
:MARKer:Y2Position?
The :MARKer:Y2Position? query returns current Y2 cursor position. This is
functionally equivalent to the obsolete :MEASure:VSTOp command/query.
If the front-panel cursors are off or are set to Binary or Hex Mode, the marker position
N O T E
values are not defined and an error is generated. Make sure to set :MARKer:MODE to
MANual or WAVeform to put the cursors in the front-panel Normal mode.
Return Format
<position><NL>
<position> ::= Y2 cursor position in NR3 format
See Also
304
(see
page
786)
on page 297), the :MARKer:Y1Position command:
"Introduction to :MARKer Commands"
":MARKer:MODE"
on page 297
":MARKer:X1Y1source"
":MARKer:X2Y2source"
":MARKer:Y1Position"
":MEASure:VSTOp"
on page 726
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
on page 296
on page 299
on page 301