Operation Status Event Register (:Operegister[:Event]) - Agilent Technologies InfiniiVision 3000 Programmer's Manual

X-series
Hide thumbs Also See for InfiniiVision 3000:
Table of Contents

Advertisement

36
Status Reporting

Operation Status Event Register (:OPERegister[:EVENt])

The Operation Status Event Register register hosts these bits:
Name
RUN bit
WAIT TRIG bit
PWR bit
MTE bit
OVLR bit
If any of these bits are set, the OPER bit (bit 7) of the Status Byte
Register is set. The Operation Status Event Register is read and cleared
with the :OPERegister[:EVENt]? query. The register output is enabled or
disabled using the mask value supplied with the OPEE command.
1068
Location
Description
bit 3
Is set whenever the instrument goes from a stop
state to a single or running state.
bit 5
Is set by the Trigger Armed Event Register and
indicates that the trigger is armed.
bit 7
Comes from the Power Event Registers.
bit 9
Comes from the Mask Test Event Registers.
bit 11
Is set whenever a 50Ω input overload occurs.
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide

Advertisement

Table of Contents
loading

Table of Contents