Operegister[:Event] (Operation Status Event Register) - Agilent Technologies InfiniiVision 2000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 2000 X-Series:
Table of Contents

Advertisement

:OPERegister[:EVENt] (Operation Status Event Register)

Query Syntax
:OPERegister[:EVENt]?
The :OPERegister[:EVENt]? query returns the integer value contained in
the Operation Status Event Register.
Table 54 Operation Status Event Register
Bit
14-12
11
10
9
8-6
5
4
3
2-0
Return Format
<value><NL>
<value> ::= integer in NR1 format.
See Also
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
(see
page
790)
Name
Description
---
---
OVLR
Overload
---
---
MTE
Mask Test Event
---
---
Wait
Wait Trig
Trig
---
---
Run
Running
---
---
"Introduction to Root (:) Commands"
":CHANnel<n>:PROTection"
When Set (1 = High = True), Indicates:
(Not used.)
A 50 input overload has occurred.
(Not used.)
A mask test event has occurred.
(Not used.)
The trigger is armed (set by the Trigger Armed
Event Register (TER)).
(Not used.)
The oscilloscope has gone from a stop state to
a single or running state.
(Not used.)
on page 150
on page 232
6
Root (:) Commands
169

Advertisement

Table of Contents
loading

Table of Contents