Manual Programming—Primary Infusion (Continued)
Continuous Infusion (Continued)
•
If the programmed continuous dose infusion is outside
the Soft Limit for that care area, an audio alert sounds
and a visual prompt appears before programming can
continue. If Yes soft key is pressed, programming
continues; if No soft key is pressed, infusion needs to
be reprogrammed.
•
If the programmed continuous dose infusion is outside
the Hard Limit for that care area, an audio alert sounds
and a visual prompt appears before programming can
continue. Infusion needs to be reprogrammed.
•
If a dose outside of the Soft Limits has been entered
and verified as correct, Message Display also shows
either "LLL" for a low dose or "
•
If a Soft Limit is overridden, G icon is displayed. When
G soft key is pressed, all applicable out-of-range limits
are listed.
Manual Programming with Guardrails Suite MX
2-32
" for a high dose.
Alaris System User Manual – with v9.33 Model 8015
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
A
Alteplase
Stroke
Dose exceeds
Yes
Guardrail limit of
No
0.81 mg/kg/h.
Proceed?
>Press Yes or No
A
Alteplase
Stroke
Dose exceeds
Reprogram
Guardrails hard
limit of 0.82 mg/kg/h.
>Press REPROGRAM
Midtown Hospital
Adult ICU
A
VTBI = 250 mL
VOLUME
AUDIO
INFUSED
ADJUST
Pump and Syringe Modules Section
G