Manual Programming—Secondary Infusion (Pump Module) (Continued)
NOTE:
It is possible to program an infusion with a rate that is
displayed with two decimal places (one-hundredth of a mL per
hour) on the PC unit for the Pump module. However, due to
space limitations on the Pump module rate display, the rate
is displayed to the nearest one-tenth of a mL per hour on the
Pump module. This value is only used for display purposes
and the Pump module is actually infusing at the more precise
rate noted on the PC unit.

Stop Secondary and Return to Primary

1. Press
CHANNEL SELECT
2.
Press
soft key.
SETUP
3.
Press
PRIMARY
4.
Close clamp on secondary administration set.
OR
Disconnect secondary administration set from upper
injection port.
5.
Press
soft key.
START
6.
To stop secondary infusion and begin infusing primary,
press Yes soft key.
Secondary infusion stops and primary infusion begins.
Main screen appears.
NOTE:
The SEC to PRI alert does not sound when the infusion is
manually ended and returned to primary.
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
Infusion (Continued)