Change Syringe And Restore Infusion - CareFusion Alaris 8015 User Manual

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

Pause Infusion (Continued)
2.
To reinitiate infusion:
Press
RESTART
OR
Press
CHANNEL SELECT
key on Main Display.

Change Syringe and Restore Infusion

1.
If syringe requires replacement:
a. Unlock security door.
b. Remove existing syringe and prepare new syringe (see
"Getting Started," "Prepare and Load Syringe and
Administration Set").
If drug and/or drug concentration is different from
previous syringe, attach and prime new administration
set.
c. Load syringe and administration set (see "Getting
Started," "Prepare and Load Syringe and
Administration Set," "Load Syringe and Administration
Set").
d. Select syringe type and size (see "Prepare Infusion,"
"Select Syringe Type and Size").
2.
To restart infusion using restored parameters, press
soft key and continue with next step.
RESTORE
OR
To start a new infusion, select drug from Drug Library and
follow steps for "Infusion Modes."
Programming
3-34
Infusion Modes (Continued)
key.
key and then press
soft
START
Alaris System User Manual – with v9.33 Model 8015
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
PCA Module
A
Guardrails Drugs
Med Surg
Morphine
A-E
Meperidine
F-J
K-O
P-T
U-Z
>Select Drug
EXIT
RESTORE
PCA Module Section