CareFusion Alaris 8015 User Manual page 252

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

Set Up PCA Dose and Continuous Infusion (Continued)
3.
To enter PCA dose, press
numeric data entry keys.
4.
To enter lockout interval, press
key and use numeric data entry keys.
5.
To enter continuous dose, press
use numeric data entry keys.
6.
To enter maximum limit, press
Yes soft key, and use numeric data entry keys.
Time (in hours) associated with
entered based on setup in system configuration.
If No is selected - then no MAX LIMITS will be available for
this infusion.
7.
To enter loading dose, press
Yes soft key and use numeric data entry keys.
Loading dose is included in VTBI but is not included in
Max Limit.
8.
Verify correct parameters and press
If the programmed parameters are outside the Soft
Limit for that care area, an audio alert sounds and a
visual prompt appears before programming can
continue. If Yes soft key is pressed, programming
continues; if No soft key is pressed, infusion needs to
be reprogrammed.
Programming
3-18
Infusion Modes (Continued)
soft key and use
PCA DOSE
LOCKOUT INTERVAL
CONT DOSE
soft key, press
MAX LIMIT
is automatically
Max Limit
soft key, press
LOAD DOSE
CONFIRM
soft
soft key and
soft key.
Alaris System User Manual – with v9.33 Model 8015
A
PCA + Continuous
Morphine
PCA
1
mg
DOSE
LOCKOUT
_ _ _
INTERVAL
minutes
CONT
_ _ _
mg/h
DOSE
MAX
None
LIMIT
LOADING
None
DOSE
[Conc]: 1 mg/mL
>Enter Lockout Interval
CHANGE
MODE
A
PCA + Continuous
Morphine
PCA
1
mg
DOSE
LOCKOUT
10
minutes
INTERVAL
CONT
2
Yes
mg/h
DOSE
MAX
> > > >
No
LIMIT
LOADING
None
DOSE
[Conc]: 1 mg/mL
>Select Max Limit Option
CHANGE
MODE
A
Guardrails Drug Setup
IMS Pre-fill 30mL
1
PCA
mg
DOSE
LOCKOUT
10
minutes
INTERVAL
2
CONT
mg/h
DOSE
MAX
20
LIMIT
_
mg/4 h
LOADING
None
DOSE
[Conc]: 1 mg/mL
>Close and Lock Door Then
Press CONFIRM
CHANGE
CONFIRM
MODE
PCA Module Section