Attach Gas Scavenging System - CareFusion Alaris 8015 User Manual

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

In the presence of high oxygen or anesthesia concentrations, it
might be necessary to connect a gas scavenging system to the
EtCO
module.
2
1.
Open gas inlet/outlet door by turning door
counterclockwise until gas outlet is clearly visible. Hold in
open position.
Gas outlet is located on lower right corner of instrument
and is marked with a gas outlet symbol (
2.
Secure gas scavenger system tubing to EtCO
firmly pushing tubing into fitting on gas outlet.
3.
Release door.
Getting Started
5-4

Attach Gas Scavenging System

Gas Scavenger System Tubing
WARNING
There is a risk of patient
cross-infection if the sampled gas is
returned to the breathing system.
).
module by
2
Alaris System User Manual – with v9.33 Model 8015
EtCO
Module Section
2