Manual Programming—Primary Infusion
The following procedures are to be used only when the drug
to be infused is listed in the Drug Library. To access the Drug
Library, a hospital-defined best-practice Data Set must be
transferred to the Alaris System and the Profile feature must be
enabled.
1.
Perform following steps (see PC unit section of this
manual, "General Setup and Operation," "Start-Up"):
a. Power on system.
b. Choose Yes or No to New Patient?
c. Confirm current Profile or select a new Profile.
d. Enter patient identifier, if required.
2.
Prepare and load syringe/administration set (see "Getting
Started").
3.
Prime (see "Getting Started").
Manual Programming with Guardrails Suite MX
2-24
Manual Programming with Guardrails Suite MX
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015