Programming with Interoperability and Guardrails Suite MX (Continued)
Pre-population of infusion parameters applies to initial and
subsequent infusions started in the Guardrails Library, as
follows:
The following primary infusions can be pre-populated for both
initial and subsequent infusion:
Continuous
Fluids
Intermittent
The following secondary infusions can be pre-populated
(Pump module only):
Intermittent
The following infusions cannot be pre-populated and require
manual programming:
Titrations
Bolus doses
Basic Infusions (including Drug Calc infusions)
All Guardrail limits that apply to manual programming also
apply to programming with interoperability. Additionally,
infusions are protected by Guardrails limits with each rate
change or titration.
The following procedures are to be used only when the drug to
be infused is listed in the Drug Library. To access the Drug
Library, a hospital-defined best-practice Data Set must be
transferred using the Guardrails Editor Software and the
Profiles feature must be enabled.
NOTE:
There is a potential for a discrepancy in precision between the
numeric values displayed on the Alaris System (PC Unit, Pump
module and Syringe module) and in the electronic medical record
(EMR)/hospital information system (HIS) due to different rounding
rules.
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
Programming with Interoperability and Guardrails Suite MX
2-59