Manual Programming—Primary Infusion (Continued)
Intermittent Infusion (Continued)
•
If Yes was selected and facility has defined a Clinical
Advisory for that drug, a message appears. To indicate
information has been noted and continue
programming, press
•
If Yes was selected to continue programming, drug
amount and diluent volume (if defined in Drug Library)
are automatically entered for selected drug.
•
If selected drug had "_ _ / _ _ mL" concentration, drug
amount and diluent volume need to be entered.
•
If selected drug is not weight-based, Not Used is
displayed in
PATIENT WEIGHT
•
If hospital/facility practice guidelines identify selected
drug as weight-based or by BSA, a prompt for a patient
weight in kilograms or BSA appears (as in illustrated
example, which reflects the use of Methotrexate).
4.
Verify correct parameters and press
confirm.
•
If the programmed total dose drug amount is outside
the Soft Limit for that care area, an audio alert sounds
and a visual prompt appears before programming can
continue. If Yes soft key is pressed, programming
continues; if No soft key is pressed, infusion needs to
be reprogrammed.
•
If the programmed total dose drug amount is outside
the Hard Limit for that care area, an audio alert sounds
and a visual prompt appears before programming can
continue. Infusion needs to be reprogrammed.
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
soft key.
CONFIRM
field.
soft key to
NEXT
A
DRUG
AMOUNT
DILUENT
VOLUME
BSA
DOSE =
[Conc]: 136.8 mg/mL
>Enter BSA
Manual Programming with Guardrails Suite MX
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
Guardrails Drug Setup
Methotrexate
6840
mg
50
mL
_ _ _
2
m
_ _ _
2
mg/m
DRUG
LIBRARY
2-41