The EtCO
module has been designed and manufactured
2
to exacting standards and should perform well within given
environmental and performance standards. There are certain
conditions under which an inaccurate measurement or the loss
of respiratory rate signal can occur.
An inaccurate EtCO
•
Incorrect disposable application or use.
•
Microstream Disposable disconnected or not securely
connected to EtCO
•
Airway connection clogged, twisted, or leaking.
•
Placement too close to electrosurgery equipment.
•
Mechanically ventilated patient breathes spontaneously.
Loss of a respiratory rate signal can occur in any of the
following situations:
•
Incorrect disposable application or use.
•
Microstream Disposable disconnected or not securely
connected to EtCO
•
Airway connection clogged, twisted, or leaking.
•
Patient not breathing.
•
Placement too close to electrosurgery equipment.
General Information
5-24
Measurement Accuracy
measurement can be caused by:
2
module.
2
module.
2
WARNINGS
•
If uncertain about measurement
accuracy, assess patient's
condition and vital signs by
alternate means, then ensure
that EtCO
correctly.
•
Leaks or internal venting of
sampled gas can affect
accuracy.
Alaris System User Manual – with v9.33 Model 8015
module is functioning
2
EtCO
Module Section
2