Prepare Syringe and Administration Set (Syringe Module) (Continued)
Prime—Using Options Menu (Continued)
Administration Set With Pressure
Sensing Disc
1.
Ensure that administration set is not connected to patient.
2.
If installed, remove pressure sensing disc from instrument.
Using a finger, apply firm downward pressure on pressure
sensing disc (not tubing) until disc snaps loose from slot in
pressure sensing disc housing.
3.
Press
key.
OPTIONS
4.
Press Prime Set with Syringe soft key.
Getting Started
2-16
CAUTION
The pressure sensing disc, if left
installed during priming, can trap air
that might not be totally expelled.
To ensure that entrapped air is
eliminated, it is recommended
that the pressure sensing disc be
removed prior to priming and the
membrane gently massaged with a
finger while priming. After priming
is completed, reinstall the pressure
sensing disc.