Programming with Interoperability—Initial Primary Infusion (Continued)
The Patient ID and, depending on the infusion type, the
following infusion parameters are pre-populated on the
Alaris System from the automated programming request:
•
Drug or Fluid name (alias or NDC)
•
Drug amount
•
Drug Amount units
•
Diluent Volume
•
Dose
•
Dosing units
•
Patient Weight or BSA (if used)
•
Rate
•
Volume to be infused (VTBI)
•
Duration
7.
Review and verify that all parameters pre-populated on the
Alaris System are correct prior to starting the infusion.
NOTE:
•
If the infusion parameters are unable to be pre-populated
on the Alaris System, error messages are displayed in the
EMR/HIS. Error messages are not displayed on the PC
unit. Consult the EMR/HIS for more information.
•
Pre-population of infusion parameters on the Alaris System
DOES NOT automatically start the infusion. As with any
infusion, all infusion parameters need to be reviewed and
confirmed by the clinician before pressing START.
8.
Start applicable infusion as described in the following
procedures:
•
Continuous Infusion
•
Fluid Infusion
•
Intermittent Infusion
Programming with Interoperability and Guardrails Suite MX
2-66
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015