Manual Programming—NO Guardrails Suite MX Protection
If Delay Options is enabled, the
.
DELAY OPTIONS
1.
(see PC unit section of this User Manual, "General Setup
and Operation," "Start-Up"):
a. Power on system.
b. Choose Yes or No to New Patient?
c. Confirm current Profile or select a new Profile.
d. Enter patient identifier, if required.
2.
Prepare and load syringe/administration set (see "Getting
Started").
3.
Prime (see "Getting Started").
4.
Start applicable infusion, as described in following
procedures:

Basic Infusion without Drug Calculation

Basic Infusion with Drug Calculation
Bolus Dose
Basic Infusion without Drug Calculation
The following procedure should be used only to set up a Basic
Infusion. To program an infusion using Guardrails Drugs, see
"Primary Infusion - With Guardrails Suite MX Protection."
1.
Press
CHANNEL SELECT
2.
Press Basic Infusion soft key.
Infusion Setup screen appears.
3.
Start applicable infusion, as described in following
procedures (see "Primary Infusion - With Guardrails Suite
MX Protection," "IV Fluid Infusion" procedure).
Rate/Volume Infusion
Volume/Duration Infusion
Programming—NO Guardrails Suite MX Protection
2-88
soft key becomes
PAUSE
key.
WARNING
When addressing or clearing an
occlusion:
Ensure that the fluid flow to the
patient is OFF to prevent
administering an unintended bolus.
An occlusion may pressurize the
infusion tubing and syringe, which
can result in an unintended bolus of
drug when the occlusion is cleared.
In order to prevent this additional
bolus, disconnect the tubing or
relieve the excess pressure through
a stopcock, if present. The clinician
should weigh the relative risks of
disconnection with the risks of an
unintended bolus of drug.
* Using a set with a pressure sensing
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART