Features and Definitions (Continued)
Operating Modes
Patient History
General Information
3-52
Features and Displays (Continued)
Four operating modes are available:
• PCA only
• Continuous Infusion
• PCA + Continuous Infusion
• loading dose only
All programming of infusions in each of four modes are completed using
Drug Library as defined by hospital-established best-practice.
PCA Module records and displays patient history for up to 24 hours, and
can be trended to following intervals: 1-hr, 2-hr, 4-hr, 8-hr, 12-hr, 24-hr.
Patient history includes following trending information:
•
total demands
•
delivered demands
•
total drug delivered
•
time and date patient history last cleared
•
average drug per hour
•
default view is 8-hr
•
total amount of drug delivered via:
◦
PCA dose
◦
Continuous Infusion
◦
loading dose
◦
Bolus Dose
Alaris System User Manual – with v9.33 Model 8015
PCA Module Section